Brand | Fluke |
Model | 572-2 |
Temperature Range | -30°C to 900°C (-22°F to 1652°F) |
Response Time | < 500 ms |
Battery Life | 8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100% duty cycle (thermometer continuously on)More |
Emissivity | Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials |
Brand | Fluke |
Model | 572-2 |
Temperature Range | -30°C to 900°C (-22°F to 1652°F) |
Response Time | < 500 ms |
Emissivity | Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials |
Backlight | Two levels, normal and extra bright for darker environments |
Hi/Low Alarms | Audible and two-color visual |
Trigger Lock | Yes |
Min/Max/Avg/Dif | Yes |
Infrared Basic Accuracy | ≥ 0°C: ± 1°C or ± 1% of the reading, whichever is greater ≥ -10°C to < 0°C: ±2°C < -10°C: ±3°C ≥ 32°F: ± 2°F or ± 1% of the reading, whichever is greater ≥ 14°F to < 32°F: ±4°F < 14°F: ±6°F |
Laser Convergence Distance | Offset dual laser, output < 1 mW |
Repeatability | ±0.5% of reading or ±0.5°C (±1°F), whichever is greater |
Optical Resolution | 60:1 (calculated at 90% energy) |
Spectral Response | 8 μm to 14 μm |
Laser Safety | IEC 60825-1 FDA Laser Class II IEC 61326-1 CE Complaint CMC # 01120009 |
Minimum Spot Size | 19 mm |
Probe Input | K-type thermocouple simultaneous display of probe and IR temperature |
K-type Thermocouple Input Temperature Range | -270°C to 1372°C (-454°F to 2501°F) |
K-type Thermocouple | 0.1°C / 0.1°F |
K-type Thermocouple Input Accuracy | < -40°C: ±(1°C + 0.2° / 1°C) ≥ -40°C: ±1% or 1°C, whichever is greater < -40°F: ±(2°F + 0.2° / 1°F) ≥ -40°F: ±1% or 2°F, whichever is greater |
K-type Thermocouple Repeatability | ±0.5% of reading or ±0.5°C (±1°F), whichever is greater |
Switchable Celsius and Fahrenheit | Yes |
Data Storage | 99 points |
Communication | USB 2.0 |
Measurement Range (Bead Probe) | -40°C to 260°C (-40°F to 500°F) |
Accuracy | ±1.1°C (±2.0°F) from 0°C to 260°C (32°F to 500°F). Typically within 1.1°C (2.0°F) from -40°C to 0°C (-40°F to 32°F) |
Cable Length | 1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector and bead termination |
Battery Life | 8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100% duty cycle (thermometer continuously on) |
Temperature Display | Dot matrix 98 x 96 pixels with function menus |
Weight | 0.322 kg (0.7099 lb) |
Display Resolution | 0.1°C / 0.1°F |
Battery Type | 2 AA batteries |
Operating Temperature | 0°C to 50°C (32°F to 122°F) |
Storage Temperature | -20°C to 60°C (-4°F to 140°F) |
Operating Humidity | 10% to 90% RH non-condensing up to 30°C (86°F) |
Operating Altitude | 2000 m above mean sea level |
Have a requirement?